In case of an inaccurate referencing of the proton NMR spectrum by the spectrometer, you have to redo the referencing for the whole NMR data set:ġ st step: Reference the TMS signal of the proton NMR spectrum: Thus, using a suitable secondary reference compound is no longer recommended! According to the IUPAC recommendation from 2001, the referencing of all other spectra will then be calculated by using the UNIFIED SCALE. If you measure a set of NMR data, firstly, the proton NMR spectrum has to be referenced to the TMS signal (0 ppm). Hint: Apply the baseline correction especially for measurements with large SW values
#HOW TO KNOW THE MHZ OF NMR SPECTRA ON MESTRENOVA ON MAC FULL#
Baseline correction: Upper menu bar: Processing ⇒ Baseline ⇒ Full Auto (Whittaker Smoother). order for the signal at the pivot point: Point the cursor into the blue area, press and shift left mouse button (LM)Ĭ) Phase correction 1 st order for the signals at the right side of the spectrum: Point the cursor into the blue area, press and shift right mouse button (RM) #HOW TO KNOW THE MHZ OF NMR SPECTRA ON MESTRENOVA ON MAC MANUAL#
Phase correction: Upper menu bar: Processing ⇒ Phase Correction ⇒ Manual CorrectionĪ) Shift the pivot point (vertical blue line) onto an intense signal at the left side of the spectrumī) Phase correction 0. 0.2 Hz)Įffect: Improves the resolution, but signal to noise gets worse 0.4 Hz) a ndĪctivate Gaussian ⇒ enter pos. Resolution enhancement: Upper menu bar: Processing ⇒ Apodization ⇒Īctivate Exponential ⇒ enter neg. valueĮffect: Improves signal to noise, but results in broader lines (try 1 Hz for carbon nmr) Line-Broadening: Upper menu bar: Processing ⇒ Apodization ⇒ activate Exponential ⇒ enter pos. Zerofilling and Linear Prediction: Upper menu bar: Processing ⇒ Zero Filling and LP. In addition, you can use the pulldown menu Processing of the upper menu bar. The middle menu bar shows icons to start most of the processing steps, mentioned further down (exception: Zerofilling). After editing all spectra, you'll get one comprehensive MNova file, which contains all NMR data you've recorded and processed before. phase and baseline correction, to get accurate integration). Nevertheless, sometimes it's necessary to redo some processing steps manually (e.g. So, after clicking on the fid or ser file you will get the full processed NMR data immediatly. There exists only one command to open and process fids. So each NMR spectrum ( 1H, 13C, gCOSY and gHSQC) gets its own page. Local Installation of MNova on your own computerĪ MNova file contains separate pages for each measurement. You will find the button to launch MNova in the APPS. #HOW TO KNOW THE MHZ OF NMR SPECTRA ON MESTRENOVA ON MAC INSTALL#
(When you visit this site for the first time, you will be asked to install the Citrix receiver) Otherwise you won't be able to load your NMR data in MNova (students in practical OC courses will find their NMR data on the N network map of NWZ, see NMR for practical OC courses)
Now you will also be able to connect to the NMR server (\\mora\nmr) to dowload your NMR data on your local disk and to process it with MNova. Use the VPN client anyconnect to connect your computer with the network of the WWU.